Resonator Description
The Sample Holders
A.1.3
There are five different sample holder sizes with 1, 2, 3, 4, and 5
mm diameter holes. Choose the smallest sample holder that
allows your sample tube to be comfortably inserted into the
holder. The sample is inserted through the threaded end. (See
Figure A-6.)
The sample rods will
not allow the use of
plastic caps on the
sample tubes.
Figure A-6
Proper way of inserting a sample tube into the
sample holder.
Spring fingers inside the sample holder gently grip the sample
and but still allow the sample to be pushed with your fingers. If
the sample slips out of the sample holder under its own weight,
the sample holder needs to be tightened. Remove the sample,
insert a screwdriver in the slot, and gently turn the disk clockwise to tighten the sample holder. (See Figure A-7.) Reinsert the
sample and check for fit.
Adjustment
Slot
Figure A-7
A-8
The end of the sample holder