Resonator Description The Sample Holders A.1.3 There are five different sample holder sizes with 1, 2, 3, 4, and 5 mm diameter holes. Choose the smallest sample holder that allows your sample tube to be comfortably inserted into the holder. The sample is inserted through the threaded end. (See Figure A-6.) The sample rods will not allow the use of plastic caps on the sample tubes. Figure A-6 Proper way of inserting a sample tube into the sample holder. Spring fingers inside the sample holder gently grip the sample and but still allow the sample to be pushed with your fingers. If the sample slips out of the sample holder under its own weight, the sample holder needs to be tightened. Remove the sample, insert a screwdriver in the slot, and gently turn the disk clockwise to tighten the sample holder. (See Figure A-7.) Reinsert the sample and check for fit. Adjustment Slot Figure A-7 A-8 The end of the sample holder