Download BROOKFIELD R/S+ RHEOMETER Operating

Transcript
Printer Error
Cause:
The printer has been selected as the output device but is not ready for
printing.
a] printer cable is not connected to the rheometer
b] paper is out, printer is not on line, other printer errors
Example of
Print Error
Message
Try again
Error #1
Printer not ready!
<OK> cont.
<ST> stop
Abort activity;
do not try again
What to Do: a] make sure printer cable is connected to the rheometer
b] make sure paper try has paper
c] check that the “ready” LED lights of the printer are on
d] remove the error cause at the printer and press OK at the rheometer
e] press ST to abort the process if you cannot find the error cause.
Zero Calibration Error
Cause:
Example
Message
An impermissibly high value was measured during the zero point calibration of the rheometer
Error #3
Zero cal. error!
Please retry cal.
<OK> cont.
<ST> stop
What to Do: a] press ST
b] ensure the measuring system has been removed from the rheometer
c] retry zero point calibration - if error occurs again, this is an indication
of a fault in the measuring instrument
Brookfield Engineering Labs., Inc.
Page 59
Manual No. M08-219