Control -menu When the Control option is selected in the main menu, a submenu is opened: 7.2.12 Device controls Charger, Trap and Flush can have either the value On or Off. The charger and the flush pump can be set On also in the chart view, when the Menu 2/3 has been selected. When measurements are made the charger and the trap should always be on. 7.2.13 Zeroing of the electrometer currents The zeroing of the electrometer currents starts when Start Electrometer Calibration has been selected. The status text Start is changed to Running… for the duration of the zeroing process. The zeroing parameters can be set in the parameter file. During the zeroing operation the system does not respond to any commands given by the user. 7.2.14 Setting the controller parameters The parameters of the PID-controller of the impactor low pressure are set by choosing Pressure PID P, I or D. A new window is opened where the values can be set. These values should not be changed. The impactor low pressure set point can be set by choosing Low Pressure Set Point. A new window is opened, where the value can be set. © Dekati Ltd. 2011, ELPI+ User Manual 42