SIBELMED W-20 User´s Manual
39
EX.- Indicates that the espiration start has not been satisfactorily,
as the extrapolated volume is superior to 5% of the FVC or 0,15
litres. The ATS recommends to be inferior to 5% of the FVC or
0,15 litres, the highest.
The Spirometry Software SIBELMED W-20 always stores all
the parameters of the different spirometric tests. Nevertheless,
the selection made here affects only the screen presentation or
in the printed reports.
PARAMETERS AND OTHERS - HEADER REPORT
Select the option PARAMETERS AND OTHERS - HEADER
REPORT into SETUP
In the printed report a three line header can be inserted with
the data selected by the user.For example: Name of the Centre,
Doctor, Address, etc.
This header will remain as long as it is not modified.
Chapter 1: Instructions of use and instalation
FP.- Indicates that the espiration has not ended satisfactorily,
as the accumulated volume of variation in the last second of the
manoeuvre is superior to 0,03 litres. The ATS recommeds less
than 0,03 litres.