Download OM-50 Manual
Transcript
6 Maintenance and Troubleshooting 6.1 Dissolved oxygen electrode maintenance ● Long-term storage 1. 2. Remove the electrode from the DO meter. 3. 4. Remove the DO tip from the electrode body. Wash the electrode well with pure (de-ionized) water, and then remove the water drops using cotton gauze. Push the socket onto the DO tip, and then store it by placing it in its original packaging and sealing it air-tight. DO tip Short socket O-ring Be careful not to damage the DO tip membrane. When using a neutral detergent for cleaning, be careful not to allow the detergent to come in contact with the membrane. When conducting air calibration after the electrode has been stored, first connect the electrode to the main unit of the DO meter and allow it to stand in the open air for two hours prior to conducting the calibration. ● Cleaning electrodes Each time a different solution is to be measured, rinse the electrode with pure (de-ionized) water, and then wipe off the water drops using clean filter paper or cotton gauze. OM-51 81