95 ml scale mark, and the other one is cut very close to the tip. The tip portions of the syringes are glued together using a hot melt glue gun. One end of the reservoir is connected to the solution line, and the other end is connected to a solution switch. This switch is used to hold the air pressure, and when it's turned on, the pressure is released. 3. Shield everything Since the solution lines behave like an antenna network, it is import to shield the exposed lines as much as possible. You may make a Faraday cage around the perfusion system using aluminum foil or plate, and make sure that the Faraday cage is grounded. If the cage is not grounded, it might introduce even more noise into the system. In addition to shielding the solution lines, you may want to shield the temperature control system as well. Although the circulating water in the temperature control system is not electrically connected to the recording chamber, it may still carry detectable noise into the recording system. Use aluminum foil to wrap all the tubings, and again, make sure that the foil is properly grounded. There is still something more to be shielded. Tubings for hydraulic micromanipulator (if used), the headstage for recording amperometric current, and the carbon electrode (except the portion that is inserted into the pipette holder) also need to be shielded. It is to reduce the noise in amperometric current. 4. Ground yourself