Download ERZ 2000 Operating Instructions

Transcript
ANNEX
...............................................................................................................................................................................................................
Temperature correction for orifice and pipe
GA10 Substance orifice
GA11 Substance pipe
Coefficients
Options
A
B
off
-
-
Linear
-
-
GOST
-
-
Steel I
12,60
0,0043
Steel II3
12,42
0,0034
Steel III
12,05
0,0035
Steel IV
10,52
0,0031
Steel V
17,00
0,0038
Steel VI
16,30
0,0116
Bronze SnBz4
17,01
0,0040
Copper E-Cu
16,13
0.0038
Gunmetal Rg9
16,13
0,0038
Brass Ms63
17,52
0,0089
Nickel
14,08
0,0028
Hastelloy C
10,87
0,0033
179
off
The relevant temperature correction is deactivated.
linear
The correction factors GA03 T-crr fact orifice and GA04 T-crr fact pipe are calculated using the linear heat
expansion coefficients GA05 Lin.expanse orifice and GA06 Lin.expanse pipe.
T Ber . fakt  1  A.lin  (Temp  20)
Material selection
The correction factors GA03 T-crr fact orifice and GA04 T-crr fact pipe are calculated using an approximation
equation
and the coefficients A and B.
T Ber. fakt  1  ( A  (Temp  20)  B  (Temp  20) 2 )  106
Coefficients are assigned automatically when a material is selected.
The permitted temperature range for the materials listed extends from -200°C to 600°C, with the exception
of copper, nickel and brass, which have an upper limit of 500°C.
...............................................................................................................................................................................................................