Download Keypoints Manual
Transcript
Temperature (deg C) If you wish to compensate for temperature, then measurement results are recalculated to what they would be o at 20 C (industry standard). Enter the mean ambient temperature here. However, you must now enter the material (and thermal expansion coefficient) for the component being measured Material As the cmm structure is made from a single material (Aluminium), it will expand and contract uniformly with temperature. It is therefore possible to compensate for the ambient temperature of the machine in the software providing that the material (and thermal expansion co-efficient for the material) is also known. Select the material of the measured component from this list. Thermal Expansion Coefficient (ppm) If you have selected ‘other’ from the material list above then you must enter the thermal expansion coefficient in ppm for the material here. Notes - 19 -