mdi Stericheck - SVP2
F.
Rinse the Tubing, Canisters, and Membranes:
Reference Document:
Please refer to the following
•
United States Pharmacopoeia (USP)/ European Pharmacopoeia
(EP)/ any other international pharmacopeia for information on
the rinse solution to be used for your application.
•
The volume of rinse solution is predetermined during the
validation procedure.
Procedure:
I.
Remove the red caps from the air vents on top of the canisters.
II.
Take out the Stericheck needle from the product container and
insert it up to its base into the rinse solution bottle.
III.
Insert the vent needle into the rinse container so as to make an
‘X’ with the Stericheck needle.
Important:
-
If using a flame, swiftly flame the needle and cool it and then insert
into the rinse solution bottle.
IV.
Set the pump speed at 55 or lower and turn on the pump. (If
using the foot control accessory, press the pedal with foot to
keep the pump running.)
V.
Invert the rinse bottle and place it into the bottle support
system.
Pic. 11: Indicative position of rinse container
VI.
UGL SVZZP2X0935E
Pump 100 ml of rinse solution into each canister. Now turn off
the pump.