Programming the Status Register System Status Groups Baseband Operation Condition Register The Baseband Operation Condition Register continuously monitors the hardware and firmware status of the signal generator. Condition registers are read only. Table 4-7 Baseband Operation Condition Register Bits Bit Description 0 Baseband 1 Busy. A 1 in this position indicates the signal generator baseband is active. 1 Baseband 1 Communicating. A 1 in this bit position indicates that the signal generator baseband generator is handling data IO. 2–14 15 Unused. This bit position is always set to 0. Always 0. Query: STATus:OPERation:BASeband:CONDition? Response: The decimal sum of the bits set to 1 Example: The decimal value 2 is returned. The decimal sum = 2 (bit 1). Baseband Operation Transition Filters (negative and positive) The Baseband Operation Transition Filters specify which types of bit state changes in the condition register set corresponding bits in the event register. Changes can be positive (0 to 1) or negative (1 to 0). Commands: STATus:OPERation:BASeband:NTRansition <value> (negative transition), or STATus:OPERation:BASeband:PTRansition <value> (positive transition), where <value> is the sum of the decimal values of the bits you want to enable. Queries: STATus:OPERation:BASeband:NTRansition? STATus:OPERation:BASeband:PTRansition? 190 Agilent Signal Generators Programming Guide