15.2 Measurement Functions (Measurement Items)
Item
Pulse width
*1:
*2:
*3:
±
Specification
• Measurement range
• Time stamp mode: 8 ns to 20 ms
• Hardware histogram mode: 8 ns to 3.2 µs
• Display resolution
• Time stamp mode: 25 ps
• Hardware histogram mode: Larger of the two values, 25 ps or (histogram X-axis span/600)
• Measurement resolution*1
• Time stamp mode and
• the polarity is
or
:
±100 ps rms±rising trigger error*2±falling trigger error*2
• the polarity is
:
±100 ps rms±rising trigger error*2±falling trigger error*2±trigger level timing error*3
• Hardware histogram mode and
or
:
• the polarity is
±(Larger of the two values, 100 ps rms or the display resolution)±rising trigger error*2
±falling trigger error*2
• the polarity is
:
±(Larger of the two values, 100 ps rms or the display resolution)±rising trigger error*2
±falling trigger error*2±trigger level timing error*3
• Accuracy*1
±Measurement resolution±(time base frequency stability x measured value)±1 ns
systematic error
• Polarity: Select from
/
/
A value measured under standard operating conditions as described in "General Specifications" after the warm-up time
has passed.
Trigger error, A input trigger error, B input trigger, rising trigger error, and falling trigger error are defined by the following
equation
X : Input amplifier noise
2
2
X +E n
En : Noise on the measured signal
S.R: Slew rate of the input signal (V/s)
S.R
The trigger level timing error is defined by the following equation.
8 mV
slew rate of the start signal
–
8 mV
slew rate of the stop signal
±
trigger level setting accuracy trigger level setting accuracy
±
slew rate of the start signa
slew rate of the stop signal
Specifications
15
IM 704310-01E
15-3