4) If the measurement is not satisfactory due to respiratory disorder or improper maneuver,
depress the START key to re-start the measurement from zero-time and repeat from procedure
2). To abort measurement, depress the CLR key.
5) Duration for MVV measurement is 12 seconds, the generally recommended test duration. To
abort measurement, depress the STOP key. The instrument however requires a measurement
of more than five seconds for analysis.
6) After measurement, the data comparison screen is
displayed. The best data is automatically selected
and marked { while the least well data is marked
_.
MVV DATA
No
~1
2
MVV
182.4
Pr
169.9
7) To repeat the MVV measurement, repeat from procedure 1).
8) Two (2) measurement data may be fully stored in memory sequenced as 1 → 2. Subsequent
measurement will overwrite the measurement data that is not marked {.
8.10.POST-BRONCHODILATOR TESTING
This instrument is capable of performing post-bronchodilator testing for each measurement.
1) For single patient testing:
a) Perform on the patient the VC, FVC, or MVV tests desired and select the best data from
the measurements made. The selected measurement shall serve as the pre-bronchodilator
data of the patient. If so required, the appropriate procedure for report printing or data
transfer should be done now.
b) Proceed in the administering of the medication to the patient. The post-bronchodilator
test is typically conducted after a fifteen-minute period.
c)
Depress the POST key to set the patient pre-bronchodilator data into memory afterwhich
the patient ID screen is displayed. The post-bronchodilator test is now ready.