Download Fukuda Sangyo ST-95 Spiroanalyzer

Transcript
OPERATION
f)
g)
4)
a)
If any problem is encountered during data
transmission, the following error message is
displayed.
Depress the ENT key to return to the ID screen.
Fix the problem and repeat the data transmission
procedure.
Entry of FRC value (FRC)
Select option 4: FRC from the UTILITY menu by
keying the numerical label (4) and the ENT key.
FRC value, as measured by other Pulmonary
Function Testing regimen, may be entered. Where
RV value and associated parameters are computed
from the entered FRC value and the VC data.
[WTX]
Communication NG
TYPE ENT KEY
[FRC]
FRC =
0.00 L
Enter the FRC value with the numeral keys.
[FRC]
b)
On complete entry, the FRC value is interpreted as
a two decimal-place number. Key entry of the
number is on a first-in last-out sequence (where
the digit is positioned to where the underline is).
To enter an FRC value of 3.02 liters, enter the "3",
"0", "2" in sequence and depress the ENT key.
c)
If an erroneous value is entered, depress the CLR key and enter correct value.
d)
If you wish to cancel the FRC entry, depress the STOP key to return to the ID screen.
5)
FRC =
3.02 L
Setting up Bronchial Challenge Testing (BASE) / Entry of Concentration during Bronchial
Challenge Testing (DOSE)
Select option 5: BASE or DOSE from the UTILITY menu by keying the numerical label (5) and
the ENT key to do Bronchial Challenge Testing.
Prior to initiation of Bronchial Challenge Test Mode, UTILITY function 5: indicates BASE.
Only during Bronchial Challenge Testing will UTILITY function 5: indicate DOSE.
After a baseline measurement data is loaded, the instrument automatically changes into
Bronchial Challenge testing mode. Once this operation is entered, the mode is kept until
a new patient ID is entered. Printing of Bronchial Challenge performance data may be
done only after saving the data set as a DOSE set data.
Page 39 of 66