Download Model FLXA21 2-Wire Analyzer
Transcript
16 nWiring Diagrams WTB10 or BA10 Terminal box *6 Case of Distributor PH201G (Style B) FLXA21 2-Wire Analyzer – CMN CMN + B D H C F A n n n n n n n n n *4 Sensor Output 1 (1-5V DC) *1 + – WTB10 or BA10 Terminal box *6 *2 (Stainless steel housing) Output 2 (1-5V DC) (*3) *5 n n n n n n n n n b *3 a HOLD *2 (Plastic housing) L N c e d f FAIL *7 Power supply 20 to 130V DC or 80 to 138V AC, 47 to 63Hz WASH Case of Distributor SDBT – + B A – + H F Output 2 (1-5V DC) Sensor Output 1 (1-5V DC) *7 L N + 1 – 2 Power supply 20 to 130V DC or 80 to 138V AC, 47 to 63Hz (*3) *1: Use a 2-wire shielded cable with an outside diameter of 6 to 12 mm. *2: Connect the analyzer to ground. (Class D ground: 100 ohm or less) The way of connecting the grounding cable varies depending on the plastic housing and stainless steel housing. In the case of the plastic housing, connect the grounding cable to the terminal of the power module inside, and in the case of the stainless steel housing, connect the grounding cable to the terminal of the housing. Use a cable with an outside diameter of 3.4 to 7 mm for the grounding line of the plastic housing. Although, on the stainless steel housing, the ground terminal symbol is (protective ground), the ground is really functional ground. *3: This line is connected to a distributor or 24V DC power supply. *4: Terminal numbers for each sensor module are shown below. *5: Two modules of the same kind of measurement/sensor type can be installed. When measuring inductive conductivity or pH/ORP with the SENCOM sensor, only one module can be installed. *6: The terminal box may be necessary depending on the sensor cable length and the distance between the analyzer and the sensor. The SENCOM sensor is to be connected directly to the analyzer without a terminal box. *7: Two outputs, output 1 and output2, of PH201G or SDBT are same signals. 11 12 14 18 13 17 19 15 16 PH Module PH 11 12 13 14 15 16 NC SC Module SC 11 12 13 17 14 16 15 ISC Module ISC 11 12 16 15 13 14 17 DO Module 18 NC DO NC 82 83 84 NC 86 SENCOM Module All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation 87 SENCOM GS 12A01A02-01E Mar. 31, 2015-00