Download Model FLXA21 2-Wire Analyzer

Transcript
16
nWiring Diagrams
WTB10 or BA10
Terminal box *6
Case of Distributor
PH201G (Style B)
FLXA21 2-Wire Analyzer
– CMN CMN
+ B
D
H
C
F
A
n n n n n n n n n *4
Sensor
Output 1 (1-5V DC)
*1
+
–
WTB10 or BA10
Terminal box *6
*2
(Stainless steel
housing)
Output 2 (1-5V DC)
(*3)
*5
n n n n n n n n n
b
*3
a
HOLD
*2
(Plastic housing)
L N
c
e
d
f
FAIL
*7
Power supply
20 to 130V DC
or
80 to 138V AC, 47 to 63Hz
WASH
Case of Distributor
SDBT
–
+ B
A
–
+ H
F
Output 2 (1-5V DC)
Sensor
Output 1 (1-5V DC)
*7
L N
+
1 –
2
Power supply
20 to 130V DC
or
80 to 138V AC, 47 to 63Hz
(*3)
*1: Use a 2-wire shielded cable with an outside diameter of 6 to 12 mm.
*2: Connect the analyzer to ground. (Class D ground: 100 ohm or less)
The way of connecting the grounding cable varies depending on the plastic housing and stainless steel
housing.
In the case of the plastic housing, connect the grounding cable to the
terminal of the power module inside,
and in the case of the stainless steel housing, connect the grounding cable to the
terminal of the housing.
Use a cable with an outside diameter of 3.4 to 7 mm for the grounding line of the plastic housing.
Although, on the stainless steel housing, the ground terminal symbol is
(protective ground), the ground is
really functional ground.
*3: This line is connected to a distributor or 24V DC power supply.
*4: Terminal numbers for each sensor module are shown below.
*5: Two modules of the same kind of measurement/sensor type can be installed. When measuring inductive
conductivity or pH/ORP with the SENCOM sensor, only one module can be installed.
*6: The terminal box may be necessary depending on the sensor cable length and the distance between the
analyzer and the sensor.
The SENCOM sensor is to be connected directly to the analyzer without a terminal box.
*7: Two outputs, output 1 and output2, of PH201G or SDBT are same signals.
11
12
14 18
13
17
19
15
16
PH Module
PH
11
12
13 14
15
16 NC
SC Module
SC
11
12
13 17
14
16
15
ISC Module
ISC
11
12
16 15
13
14
17
DO Module
18 NC
DO
NC 82
83 84 NC 86
SENCOM Module
All Rights Reserved. Copyright © 2010, Yokogawa Electric Corporation
87
SENCOM
GS 12A01A02-01E Mar. 31, 2015-00