Download Instructions for
Transcript
. This quick tutorial provides step-by-step instructions that will take you through the basic operating procedures and give you a general understanding of how the IAQ Monitor is used. If you have used the IAQ Monitor before, or if you are comfortable using instrumentation, this will probably give you the information you need to get started. Refer to the remainder of this reference manual for detailed instructions. This tutorial is very straightforward and will be easy to follow even if you have never used the IAQ Monitor and are not familiar with technical instrumentation. It will help get you acquainted with the operating procedures you will be using. NOTE: The CO2, temperature and relative humidity sensors are built into the IAQ Monitor. See "Connecting and Calibrating Your Sensors" in Chapter 4 for information on connecting the toxic gas and linear sensors. Turning the IAQ Monitor On and Off 1. Press the ON/OFF button. The IAQ Monitor will turn on and automatically begin displaying the present readings. 2. Press the ON/OFF button again. appear: The following screen will Press RECORD To Turn Off ON/OFF to Continue 3. Press RECORD. The unit will turn off. NOTE: In order to turn off the IAQ Monitor it CANNOT be recording. Calibrating the CO2 Sensor 6 1. Unscrew and remove the toxic sensor cap (see the IAQ Monitor instrument diagram in the beginning of Chapter 3). 2. Slide the calibration adapter over the wand until the "O" ring in the bottom of the adapter slides over the handle of the wand.