The PALM Probe displays the irradiance values in watts or milliwatts per
square centimeter (W/cm2 or mW/cm2) on the LCD. The user has to
carefully note the position of the decimal point in the reading and
whether or not the display is showing W/cm2 or mW/cm2 on the display.
The floating decimal point on the numerical display allows values as low
as 0.001 mW/cm2 or 1 µW/cm2.
For help in understanding prefixes and symbols ‘milli’ (m) and ‘micro’
(µ) please see the section on metric system basics in the appendix.
The PALM Probe is designed to measure UV Irradiance values up to 10
watts/cm2, which is higher than UV irradiance values of most current UV
sources.
Sample and Hold
When collecting irradiance data with the PALM Probe, the display will
change as the Input Aperture changes position under the UV source. As
the Input Aperture moves into the focused peak irradiance area of the UV
source, the reading will increase.
UV housing and
reflected light
Profile of UV
irradiance (y-axis) vs.
time, obtained with
EIT PowerMAPTM
Figure 8-Irradiance variations