Download QC-CALC User`s Manual
Transcript
Report - Reference Section This information tells QC-CALC to read the data and calculate the Average and Range between records 1, 4, 7, 10, & 13. Notice these records contain part #1 information only. This sequence of inspection is critical. The Multi-Part Gage Report is designed to show gage variation based on repeated measurement of a group of similar parts such as 10 parts measured 2 times. Gage variation is calculated from the range of values between the trials (number of times the parts were measured). This report shows fixture and gage errors if the parts are removed and reinstalled on the gage. This test is better than the Single-Part Sigma Report because it tests the gage's performance over a range of different parts, but more work is required to measure each part several times. It is assumed the parts do not change size over the course of the testing period. Record 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 Part 1 2 3 1 2 3 1 2 3 1 2 3 1 2 3 Unique Part count and Total Parts Measured were added to the Trial count so you can understand the conditions of the gage study. The Single-Part Gage Report only requires the Trial count. Repeatability Minimum A gage should not contribute more than 10% to the measurement error. Any value less than or equal to 10% is considered good. The P/T% column is the P/T Ratio multiplied by 100 with an added percent sign. This report does not replace the more exhaustive testing of a complete Gage Repeatability & Reproducibility QC-CALC Real-Time® Page - 44 Trial 1 1 1 2 2 2 3 3 3 4 4 4 5 5 5