A6T/A6/A5 Portable Ultrasonic Diagnostic System z Simpson (Simpson) Measurement items: LVAMd, LVLd, LVAPD, LVAMs, LVLs, LVAPs; Calculation items: EDV, ESV, SV, CO, EF, SI, CI Use the formula: ( LVLd × 4 × LVAMd + 2 × LVAPd + LVAMd × LVAPd EDV= 9 ( LVLs × 4 × LVAMs + 2 × LVAPs + LVAMs × LVAPs ESV= 9 ) ) Measurement method: ↓ ↓ 1. In B mode freeze the image, press the 【Calc】 key to pop up the Meas & Calc menu, move the trackball, use the cursor to select “Cardiac”, press the【SET】key to go to the submenu, move the trackball to select “LV” , press the 【SET】key to go to the submenu, move the trackball to select “Simpson”, press the【SET】 key to go to measurement state. 2. Measure LVAMd, see “Area/Cir measurement (Ellipse)” for the specific measurement method. 3. Measure LVLd, see “Distance Measurement” for the specific method. ↓ 4. Measure LVAPD; see “Area/Cir measurement (Ellipse)” for the specific measurement method. 5. Unfreeze the image, rescan the image, and freeze the image. 6. Measure LVAMs, see “Area/Cir measurement (Ellipse)” for the specific measurement method. 7. Measure LVLs, see “Distance Measurement” for the specific method. 4701-0040-01A 3-75