FREQUENTLY ASKED QUESTIONS
FAQs listed in Table 17 provide information about the
internal operation of WaveMaster, and answers to the
most commonly-encountered problems.
Table 17. Frequently Asked Questions
QUESTION
RESPONSE
How does it work?
The WaveMaster is a high diffraction order spectrometer. Wavelength is determined by careful measurement of the separation
between the diffracted orders. Much of the accuracy and simplicity
in use is achieved through software-implemented algorithms running
in the digital signal processor and conventional microprocessor in
the instrument.
Why won't it power up?
If WaveMaster is connected to the mains power unit and the yellow
pulse received light does not light, then the rear panel connector may
not be making a good connection and should be re-inserted. It is also
possible that the power supply has failed. This should give 12V DC
with the centre of the connector positive.
Why does it power up but
powers off very shortly
afterwards?
This does happen very occasionally. Restarting the instrument a
second time should clear the problem.
A persistent failure to start indicates a serious internal problem. The
message that appears briefly on the display provides some information about the cause of the failure. The most likely cause is that the
internal optics have become misaligned (probably through undue
shock) or the spectral calibration source has failed. Contact Coherent
with as much detail about the display message as possible (refer to
“Obtaining Service” on page 49).