Click button Place pressure plate to shafts and automatically configure the ZIF socket
actuation unit (<Insert>) and follow the instruction shown by software. This will
automatically configure the ZIF socket actuation unit for detected programming module.
Check if the pressure plate can fully open the ZIF socket, use buttons Open ZIF socket
(<PageDown>) and Close ZIF socket (<PageUp>) to open and close the ZIF.
If there is necessary to adjust the pressure plate follows the steps below:
Move the pressure plate pressure plate upwards or downwards until it will be a few
millimeters above the ZIF socket.
To move pressure plate click (for one step) or press and hold (for continuous moving)
button Pressure plate down (<Down>) or Pressure plate up (<Up>).
For faster moving press and hold Shift key both with button (or shortcut key) for moving
plate.
To leveling the pressure plate use the buttons Upper Up, Upper Down, Lower Up or Lower
Down.
Adjust elevation of the pressure plate so that it'll touch the ZIF socket, but not make a
pressure to it (preset closed position). This can be checked by
looking the moment when the pressure plate touches the ZIF socket or
feeling in fingers, get hold of the ZIF socket by fingers and try to move with the socket,
simultaneously move the pressure plate downwards; if you’ll feel that you can not move
with the socket, then the pressure plate touches the ZIF socket - now make one or two
upward steps with the pressure plate
Click the button Set current position of pressure plate as closed ZIF position
(<Ctrl+PageUp>). A position for closed ZIF socket will be remembered.
Adjust the elevation of pressure plate so that it'll fully open the ZIF socket. Be also aware of
leveling of pressure plate so, that equal pressure is brought to both ZIF's side.
Click the button Set current position of pressure plate as opened ZIF position
(<Ctrl+PageDown>). A position for opened ZIF socket will be remembered.
Click the button Save configuration of ZIF socket actuation unit (<Ctrl+S>) to store ZIF
socket actuation unit configuration.
From this point the ZIF socket actuation unit is configured and can work.
Command line parameters
Program PG4UWMC supports following command line parameters:
/prj:<file_name>
Loads project file. Parameter <file_name> means full or relative project file path and name.
There is also available to make Load project operation from command line by entering project
file name without prefix /prj:
Example:
pg4uwmc.exe c:\projects\myproject.eprj
Makes load project file "c:\projects\myproject.eprj".
/autoconnectsites
The command forces PG4UWMC to connect programmer Sites (start control program PG4UW
for each Site) during start of PG4UWMC, for all Site-s that were used, when PG4UWMC was
recently closed. There is also equivalent option "Auto-connect Sites" available in "Settings"
dialog of PG4UWMC.
143