Download ManoScan User Manual

Transcript
TM
ManoScan
User Manual
_____________________________________________________________________________________________________________
ManoScan V (A400 module): The A400 module should be connected to the computer and power
turned on before switching on the computer. The connections from A400 module to the computer is
made using the custom extender cable supplied with the system.
A400 module provides for three types of video input connectors, any one of which can be used at a
time to feed video data (from an external video source such as fluoroscopy or endoscopy equipment).
The three types of connectors and the corresponding input video format supported by them are:
a) DVI: Supports color, monochrome, or gray scale analog component video. The video input can be
of non-standard high resolution and frequency (typically, from fluoroscopy equipment)
b) BNC: Supports monochrome or gray scale analog video. The video must be of standard television
specifications: PAL or NTSC.
c) S-Video: Supports monochrome or gray scale analog composite video. The video must be of
standard television specifications: PAL or NTSC (typically, from endoscopy equipment)
ManoScanTM V can also be used to acquire video from USB based devices (e.g. webcams). The A400
module is not required for such acquisition, if the USB input is directly connected to the computer,
the appropriate video input device can be selected within the software.
2.7 Warnings
1. Ensure all personnel responsible for operating the ManoScanTM system are properly trained prior
to use.
2. Please follow the on-screen instructions and those in Section ‎3.3 to calibrate probe and verify its
proper function before each clinical use.
3. All accessory equipment connected to the analog and digital interfaces must be certified to the
respective IEC standards (i.e. IEC 60950 for data processing equipment and IEC 60601-1 for
medical equipment.) Furthermore all configurations shall comply with the system standard
IEC 60601-1-1. Any person that connects additional equipment to the signal input part or signal
output part configures a medical system, and is therefore, responsible that the system complies
with the requirements of the system standard IEC 60601-1-1. If in doubt, consult the technical
services department or your local representative.
4. When the disposable sheath is used, it must be secured to the probe to prevent accidental
aspiration. The clinician must ensure both the catheter and the sheath are removed from the
patient upon extubation. While not likely, it is possible for the sheath to slip off of the catheter
during removal.
5. The disposable sheath should not be used for proximal gut studies or pressure with impedance
studies.
6. DO‎ NOT‎ immerse‎ the‎ probe‎ in‎ liquids‎ beyond‎ the‎ “Do‎ Not‎ Immerse”‎ label on the probe.
Immersion of the connectors or‎vent‎tube‎located‎at‎the‎“Y”,‎near‎the‎proximal‎end‎of‎the‎probe,‎
can damage or destroy the probe.
7. This system contains a type BF applied part. It is not suitable for use in the presence of a
flammable anaesthetic mixture with air or with oxygen or nitrous oxide.
This document contains proprietary information. Source: Given Imaging
Copyright 2013, Given Imaging. All rights Reserved
11