Download 6100A/6101A - Alternative verification methods
Transcript
In Table 6. it is assumed the 2A range is chosen for comparison with each voltage range. Current Range: 2A Current setting: 1.6A V to reference A40 / A40A Shunt Voltage Range Setting Low Flicker value High Flicker value Frequency: 60Hz Reversed High Flicker value 6100A Low specification Calculated Flicker value error (m°) (Frequency <70Hz) 2A 16V 0.2V 0.003° 2A 33V 1.2V 0.003° 2A 78V 8V 0.003° 2A 168V 0.2A 0.003° 2A 336V 0.4A 0.003° 2A 1008V 0.8A 0.003° Table. 6 Phase measurements to single current range Phase angle can be verified at other frequencies by choosing the same appropriate harmonic for voltage and current. The voltage and current channels should be set to Harmonic mode (user manual section 4-22). Set up the required harmonic bearing in mind the amplitude limitation for harmonics (see user manual section 1-8). For harmonic frequencies < 2850Hz amplitudes up to 30% of range may be set. The maximum amplitude allowed decreases with frequency to 20% at 6kHz. Set voltage and current 1st harmonic amplitude to zero to remove the fundamental frequencies from the outputs. See user manual section 1-28 for phase angle accuracy specification of signals > 69Hz or where either the voltage or current output is less than 40% of range. - 10 -