EFOSC2 USER’S MANUAL - 4.0
4.3.28
EFOSC spec cal Arcs
LSO-MAN-ESO-36100-0004
This template is used for taking arc line spectra for wavelength calibration in spectroscopy. Helium
(LAMP1) and Argon (LAMP 2) lamps are available. It acquires arc line spectra through the desired
Filter (usually Free), Starplate and Grism. The user can control the exposure time of the 2 lamps
independently. The Calibration Unit is automatically moved IN and OUT at the beginning and at
the end of the execution respectively. Once the Calibration Unit is moved in, the lamps are turned
on sequentially, and the shutter is opened for the desired duration for each of the lamp. There is
a lamp preheating of a few seconds before each exposure. The CCD configuration can be changed,
such as the binning, readout speed, and windowing.
For a discussion of wavelength calibration issues, please see § 3.5.3. A table of the recommended
exposure times are available online at
http://www.eso.org/sci/facilities/lasilla/instruments/efosc/inst/Perf_HeAr.html
These values may change from time to time due to ageing of the lamps. The user may want to
experiment with the exposure times during afternoon calibration to optimise the line intensities.
Parameter
Default
Description
Filter
Starplate
Grism
He lamp exposure time
Ar lamp exposure time
CCD readout speed
CCD X binning
CCD Y binning
CCD windowing flag
First column of window
First row of window
Number of columns
Number of rows
Number of exposures
Free
NODEFAULT
NODEFAULT
NODEFAULT
NODEFAULT
normal
2
2
F
1
1
2048
2048
1
User specified filter
User specified slit
User specified grism
Exposure time for He lamp (sec)
Exposure time for Ar lamp (sec)
CCD read-out speed
CCD binning factor in X direction
CCD binning factor in Y direction
F=no CCD window, T=CCD windowed
If ST=T, first column of window
If ST=T, first row of window
Number of CCD columns in window
Number of CCD rows in window
Number of exposures