Download MuSR Manual
Transcript
Figure 7: Range curve for the Blue Variox Cryostat. The range curves are different for each cryostat and those which are significantly different from the Variox curve shown here are included in the sections for each cryostat. 2.5 Magnets The MuSR spectrometer has three sorts of magnets acting on the sample position: zero field compensation coils, calibration ‘T20’ coils, and the Helmholtz coils that act as the main magnet. They are set to act individually through the SECI program. 2.5.1 Main magnet The main magnet can be used to provide up to 2500G in longitudinal mode and around 600G in transverse mode (limited by the frequency response). To switch on the main magnet use the SECI command lf0. SECI should write “Danfysik power supply has been selected”. After this commands of the form setmag nn, where nn is the desired field, can be used to set the field. 2.5.2 Calibration coils The calibration coils are used when the spectrometer is in longitudinal mode to provide a small transverse to calibrate the difference in count rates in the forward and backward detectors. They hang from flexible hooks next to the cryostat space and have a power cable leading from the downstream wall of the spectrometer area. When inserting them, before putting the cryostat into the spectrometer, check that they are facing in the correct direction – shown by the Out labels painted on the coils.