VT Plus HF Operators Manual Ending the Stacked Volume Test To end the Stacked Volume test: 1. Either click the STOP button or let the Stacked Volume test end following the number of requested breaths. 2. Click the Print button to print a final report for the Stacked Volume test. 3. Click the New Test button to start a new Stacked Volume test. Serial Communications The Analyzer communicates with a PC through a serial (COM) port. If the Analyzer is connected to a computer and turned on when the software is started, the software automatically detects which COM port the Analyzer is on and then initiates communications between the Analyzer and your PC. Selecting COM Port While running VT for Windows the user may select Com Port from the Communications menu to change communications settings. Selecting the Serial Communications Mode The Analyzer has four modes for data transmission via the serial port, available from the Communications menu. The currently selected mode shows a check mark next to its name. Test Parameter –the parameter that was chosen for running this test; this value does not change during the test. Wave Form – starts the continuous transmission of waveform data from the Analyzer; flow, volume, airway pressure, auxiliary pressure, and differential pressure data are received at a 50 Hz rate. Both – starts the automatic transmission of both waveform data and parameter data, as described above. Parameters – start the automatic transmission of parameters that are calculated at the end of a breath; these parameters are calculated on a breath-to-breath basis. Note No parameters are calculated for a breath if a zero occurs during that breath. None – selects no data transmission; this stops all automatic data transmissions from the Analyzer. 4-20