SECTION 5
CALIBRATION
5.1
REQUIRED EQUIPMENT
The Elgar PV simulator supports fully automated calibration, which requires the
following equipment, software and associated documentation:

Agilent model 34401A or Keithley 2000 Digital Multimeter

TerraSAS Calibration Suitcase, Elgar p/n 5609175-25

TerraSAS Calibration software v2.0.0.2 or above, Elgar p/n 5609174-XX

TerraSAS Calibration User Manual, Elgar p/n M609174-01
5.2
OPERATION
The calibration suitcase has an internal, calibrated precision shunt. It connects to the
simulator output, the digital multimeter and the AUX I/O connector on the rear panel.
All cables and accessories are included in the suitcase.
After installing and running the software, the user is guided through a simple procedure,
which takes about five minutes to complete.
5.3
CALIBRATION INTERVAL
The recommended calibration interval for all Elgar PV Simulators is one year.