Download 2.29Mb - AMETEK Programmable Power
Transcript
SECTION 5 CALIBRATION 5.1 REQUIRED EQUIPMENT The Elgar PV simulator supports fully automated calibration, which requires the following equipment, software and associated documentation: Agilent model 34401A or Keithley 2000 Digital Multimeter TerraSAS Calibration Suitcase, Elgar p/n 5609175-25 TerraSAS Calibration software v2.0.0.2 or above, Elgar p/n 5609174-XX TerraSAS Calibration User Manual, Elgar p/n M609174-01 5.2 OPERATION The calibration suitcase has an internal, calibrated precision shunt. It connects to the simulator output, the digital multimeter and the AUX I/O connector on the rear panel. All cables and accessories are included in the suitcase. After installing and running the software, the user is guided through a simple procedure, which takes about five minutes to complete. 5.3 CALIBRATION INTERVAL The recommended calibration interval for all Elgar PV Simulators is one year. 25