Download "user manual"
Transcript
RM100 User’s Manual
Command
Parameters
:SAMPle
:SAVE
:SENSe
:NULL
:STATe
Equivalent to pressing the Save key.
Transfers the measurements stored
in the data buffer to non-volatile
memory.
{ON|OFF|AUTO}
:SENSe
:NULL
:STATe?
:SENSe
:NULL
:VALUe
Equivalent to pressing the Null key
(ON) to initiate the Null function or
the Shift-Null key (OFF) to clear the
offset field or Shift-Offset key
(AUTO) to initiate the Auto Null
function.
Returns the null state.
OFF, ON or AUTO
{<value>|MIN|MAX}
:SENSe
:NULL
:VALUe?
:SENSe
:RANGe
Description
Sets the value of the offset field to
<value> which can range from 99999 to +99999 nT.
Returns the current value of the
offset field in nT.
{<value>MIN|MAX}
Sets the range of the displayed
difference field and the analog
output.
<value> represents the maximum
expected signal level in uT and must
be between 0.1 and 100. The
RM100 selects the appropriate range
for the requested value.
MIN=0.1 µT; MAX=100 µT
:SENSe
:RANGe?
:SENSe
:SMOothing
:POINts
Returns the current range of the
difference field display and the
analog output.
<value>
<value> must be between 1 and 100.
The RM100 selects 1, 3, 10, 50, or
100 points based on <value>.
:SENSe
:SMOothing
:POINts?
:SENSe
:UNITs
June 2004
Sets the number of points in the
smoothing filter running average.
Returns the number of points being
used by the smoothing filter running
average.
{uT|nT|mG}
Sets the units of the displayed
difference field.
:SENSe
:UNITs?
Returns the current units of the
displayed difference field.
:SENSe
:TEMPerature?
Returns the internal temperature of
the electronics unit in degrees C.
Resolution is 0.1 degrees.
28
MEDA, Inc.