To use EXCORR, the XCorr icon must be present in the hardware configuration. Double click on this icon, then uncheck the box for Rhodamine Quantum Counter. The Excitation Correction check box must be enabled in the dialog from the hardware configuration menu under PHB1 Configure Corrections, and the proper correction lookup table must be selected. In most cases, this is the lookup table provided by PTI. Note. The EXCORR curve provided by PTI is fully valid at the 5-nm slit at which it was measured. If narrower slit widths are used, the EXCORR curve may need to be measured again at the slit width required for your measurement. Please call PTI for details. In addition, Real Time Correction must be enabled in the Additional Acquisition Setup Controls dialog (opened by clicking the More button in an experiment dialog). The gain should be set correctly as well. This is set by clicking the Reference Source Gain button to display the XCorr Gain Control dialog. Enter the Excitation Wavelength into the edit box, and press the Goto button. The excitation wavelength is generally the excitation wavelength that gives the largest emission value (for scans), or the wavelength for the timebased experiment. Move the slider control until the Signal measures ~1 volt. Press Ok to save the changes. The program will display the corrected curve, the raw data, and the raw photodiode output upon acquiring data. During scanning, EXCORR will correct for fluctuations in illumination output due to the lamp power envelope and variations in the arc itself Emission Correction (EMCORR) The emission correction is much simpler than excitation correction. The data trace, which may already be excitation corrected, is multiplied by the correction file provided by PTI. This correction is intended to compensate in wavelength dependant variations in the system on the emission channel. Each emission channel grating has its own EMCORR file. The EMCORR file is generated by comparison of the emission channel response to the spectrum of a NIST traceable tungsten light. Note. If the grating, PMT voltage, or PMT is changed, the Emission Correction should be re-calibrated for best results. The emission channels will also undergo changes as they age, and the Emission Correction should also be re-evaluated periodically. Please contact PTI for details. To use the EMCORR file in real-time, the Emission Correction check box must be enabled in the dialog from the hardware configuration menu under PHB1 Configure Corrections, and the proper correction lookup table must be selected. This lookup table is provided by PTI, generally with the names emcorri (for channel 1) or emcorrii (for channel 2). These must be matched with the proper channel, and are properly configured by PTI personnel during installation. Next, turn on the Real Time Correction in the Additional Acquisition Setup Controls dialog (opened by clicking the More button in an experiment dialog) and selecting Emission. The raw and corrected traces will be displayed. 194