Download Notes - Artisan Technology Group
Transcript
Calibration Isolation Status Register Bit Address Base + 18816 READ The Calibration Isolation Status Register provides a non-volatile record of the isolation status of each channel at the last calibration. This register is non-volatile. 15 Ch 16 WRITE Iso Cal Stat 14 13 12 11 10 09 08 07 06 05 04 03 02 01 00 Ch 15 Iso Cal Stat Ch 14 Iso Cal Stat Ch 13 Iso Cal Stat Ch 12 Iso Cal Stat Ch 11 Iso Cal Stat Ch 10 Iso Cal Stat Ch 09 Iso Cal Stat Ch 08 Iso Cal Stat Ch 07 Iso Cal Stat Ch 06 Iso Cal Stat Ch 05 Iso Cal Stat Ch 04 Iso Cal Stat Ch 03 Iso Cal Stat Ch 02 Iso Cal Stat Ch 01 Iso Cal Stat Ch XX Iso Cal Stat: This field is set to a one (1) when the last calibration of the channel was performed with the channel configured as non-isolated. This field is set to a zero (0) when the channel was last calibrated in the isolated configuration. Notes This register is not set or used by the hardware. The HP E1418A SCPI driver, if used to calibrate the module, automatically sets the bits in this register. If you use register programming to calibrate this module, you may wish to use this register to record the calibration status. The Voltage Calibration Status Register, the Current Calibration Status Register, the Calibration Isolation Status Register, and the Calibration Card Configuration Register are used together to maintain the overall calibration status of the module. Changes in the values of these fields indicate a change in the module configuration and may indicate the need to recalibrate the module. 146 HP E1418A Register-Based Programming Appendix B Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com