Download User Manual

Transcript
M3B Vital Signs Monitor User Manual
CO2 Monitoring
AWRR ALM HI: when parameter value exceeds this limit, there will be alarm for exceeding
upper limit.
Default:
Adult:
30 rpm
Pediatric:
30 rpm
Neonatal:
100 rpm
AWRR ALM LO: when parameter value is smaller than the limit, there will be alarm for
exceeding lower limit.
Default:
Adult:
8 rpm
Pediatric:
8 rpm
Neonatal:
30 rpm
FiCO2 ALM HI: when parameter value exceeds this limit, there will be alarm for exceeding upper
limit.
Default:
Adult:
4 mmHg
Pediatric:
4 mmHg
Neonatal:
4 mmHg
WAVE SCALE: LOW/HIGH
Default:
LOW
WAVEFORM: FILLED/UNFILLED
Default:
UNFILLED
BARO PRESS:
400 ~ 850mmHg
Default:
760mmHg
O2 COMPENS:
0 ~ 100%
Default:
16 %
ANE AGENT:
0 ~ 20%
Default:
0.0%
BALAN GAS:
ROOM AIR/N2O/HELIUM
Default:
ROOM AIR
APNEA ALM:
Selections are 10s to 40s
Default:
20s
9.4 Alarm Information and Prompt
Tables below describe the possible physiological alarms, technical alarms and prompt messages
occurring during CO2 measurement.
Physiological alarms:
- 65 -