5.6 CHANnel Group
:CHANnel<x>[:VOLTage]:SCHannel<x>:
LABel
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:BVALue
Syntax
Syntax
Function
Sets or queries the label string of the specified
sub channel for the specified 16-CH voltage input
module.
:CHANnel<x>[:VOLTage]:SCHannel<x>:
LABel {<String>}
:CHANnel<x>[:VOLTage]:SCHannel<x>:
LABel?
CHANnel<x>'s <x> = 1 to 16
SCHannel
<x>'s <x> = 1 to 16
(the upper limit varies
depending on the module
type)
<String> = Up to 16 characters
Function
Example
Sets or queries the offset value B that is used
during linear scaling for an installed 16-CH
voltage input module.
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:BVALue {<NRf>}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:BVALue?
<x1> = 1, 3, 5, 7, 9, 11, 13, 15
<x2> = 1 to 16
<NRf> = –9.9999E+30 to –1E–30, 0,
1E–30 to 9.9999E+30
:CHANNEL3:VOLTAGE:SCHANNEL1:LSCALE:
BVALUE -25
:CHANNEL3:VOLTAGE:SCHANNEL1:LSCALE:
BVALUE?
:CHANNEL3:VOLTAGE:SCHANNEL1:
-> LSCALE:BVALUE -25.0000E+00
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale?
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:DISPlaytype?
Function
Queries all the linear scaling settings of an
installed 16-CH voltage input module.
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale?
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:AVALue
Function
Syntax
Example
Sets or queries the scaling coefficient A that is
used during linear scaling for an installed 16-CH
voltage input module.
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:AVALue {<NRf>}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:AVALue?
<x1> = 1, 3, 5, 7, 9, 11, 13, 15
<x2> = 1 to 16
<NRf> = –9.9999E+30 to –1E–30,
1E–30 to 9.9999E+30
:CHANNEL3:VOLTAGE:SCHANNEL1:LSCALE:
AVALUE 25
:CHANNEL3:VOLTAGE:SCHANNEL1:LSCALE:
AVALUE?
:CHANNEL3:VOLTAGE:SCHANNEL1:
-> LSCALE:AVALUE 25.0000E+00
Description The coefficient cannot be set to 0.
Function
Syntax
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:DISPlaytype:DECimalnum
Function
Syntax
Example
5-60
Queries all the linear-scaling display-mode
settings of an installed 16-CH voltage input
module.
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:DISPlaytype?
Sets or queries the decimal place that is used
during linear scaling when the display mode is
set to Float for an installed 16-CH voltage input
module.
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:DISPlaytype:
DECimalnum {<NRf>|AUTO}
:CHANnel<x1>[:VOLTage]:SCHannel<x2>:
LSCale:DISPlaytype:DECimalnum?
<x1> = 1, 3, 5, 7, 9, 11, 13, 15
<x2> = 1 to 16
<NRf> = 0 to 3
:CHANNEL3:VOLTAGE:SCHANNEL1:LSCALE:
DISPLAYTYPE:DECIMALNUM AUTO
:CHANNEL3:VOLTAGE:SCHANNEL1:LSCALE:
DISPLAYTYPE:DECIMALNUM?
-> :CHANNEL1:VOLTAGE:SCHANNEL1:
LSCALE:DISPLAYTYPE:
DECIMALNUM AUTO