Download FlexTest SE Controller Service
Transcript
Calibrating an LVDT Task 5 Set the zero and offset Establishing zero requires the actuator to be set at mid-displacement when you calibrate the LVDT for equal amounts of actuator extension and retraction. Suppose you have an actuator with ±10 cm displacement—which actually has a 20 cm displacement. Setting zero at mid-displacement produces a displacement of ±10 cm (this is the most common approach). However, you can set zero anywhere within the full scale of the sensor, such as with the actuator fully extended or retracted to produce a displacement range of 0 mm - 20 mm. Before beginning, ensure the following are true: • The Fine Zero adjustment is set to 0. Stand-Alone—Path: Setup > Disp > Calibration > Fine Zero Automated—Adjust Fine Zero on the Calibration tab. • The Manual Offset is set to 0. Stand-Alone—Path: Setup > Disp > Offset/Zero > Manual Offset Automated—Adjust Manual Offset on the Offset/Zero tab. • The Electrical Zero adjustment is set to 0. Stand-Alone—Path: Setup > Disp > Offset/Zero > Electrical Zero Automated—Adjust Electrical Zero on the Offset/Zero tab 1. Mount the displacement measuring device. Mount the measuring device to measure the distance between the end of the actuator’s piston rod and a stationary point such as the actuator’s upper endcap. There are a variety of ways to measure actuator displacement: Dial Indicator • Tape Measure/Ruler • Optical Detector Calibration Model 493.02 Controller Service • 111