Download FlexTest SE Controller Service
Transcript
Calibrating a Force Sensor Conditioner Output Voltage Gain = ----------------------------------------------------------------------------------------------------------Excitation Voltage × Compression mV/V Where: Conditioner Output Voltage is typically 10 Vdc. Excitation Voltage is specified using the Excitation entry box on the Calibration submenu or tab. Compression is specified using the Negative Compression or Positive Compressive entry box on the Calibration submenu or tab. mv/v Positive Tension Calibration Use the following procedure if your force transducer is set up so that a positive output represents actuator retraction (tension). 1. Select mv/v Pos Tension for your calibration type. Stand-Alone— Path: Setup > Force > Calibration > Cal Type > mv/v Pos Tension Automated—Select mv/v Pos Tension for Cal Type on the Calibration tab of the Inputs panel. 2. From the Calibration Data sheet for your force transducer enter the following values on the Calibration submenu or tab: Stand-Alone: A. Enter the full scale force values in the Minimum and Maximum entry boxes. B. Enter the tension sensitivity value (+mV/v) in the Pos Tension entry box. C. Enter the compression sensitivity value (-mv/V) in the Neg Compression entry box. Calibration D. Enter the calibration excitation value (Vdc) in the Excitation entry box. 154 Model 493.02 Controller Service