SAFETY AND WARNINGS
The following safety precautions apply to both operating and maintenance personnel and must be observed during all
phases of operation, service, and repair of this instrument. Before applying power, follow the installation instructions and
become familiar with the operating instructions for this instrument.
Digital Storage USB and design of IEC1010 with wavelet table 1 - safety norms, over-voltage electric measurement
category III - 1000V CAT, Pollution protection level: 1.
Caution: The rules listed below should be carefully followed for safe operation.
1.
NEVER apply voltage or current to the meter that exceeds the specified maximum for the function selected.
Function
V DC
V AC
mA AC/ DC
A AC/ DC

Diode
Capacitance
Input Jacks
V/,COM
V/,COM
400mA,COM
20A,COM
V/,COM
V/,COM
V/,COM
Maximum Input
1000Vp, within 10 seconds
700V AC RMS, within 10 seconds
500mA DC/AC RMS, Fused
20A DC/AC RMS, within 30 seconds with a 15 minutes cool down period
250V DC + AC peak, within 10 seconds
250V DC + AC peak, within 10 seconds
250V DC + AC peak, within 10 seconds