Download User Manual
Transcript
C RY S TA L GUARD (Ver-4.5) 2.2 Site Requirements CRYSTAL is designed to work in a normal environment. The unit should be placed away from 1 Processes producing electricity, fumes, dust, gasses or radiation (e.g. silicon or halon contamination from photo copying or fire protection equipment) 2 Electrical meters or switchgears 3 Water pipe, gas meters, pipes or appliances 4 Dampness 5 Vibration 2.3 General Ensure a good power point for the exchange with a 5 pin socket & switch. If possible, the customer should be asked to provide a “ clean” power supply direct from the power distribution board. CRYSTAL is set to work at 220VAC, 50Hz mains supply. WARNING: ANTI-STATIC PRECAUTIONS MUST BE TAKEN WHEN WORKING WITH THE CARDS OR COMPONENTS OF CRYSTAL EPAX. CAUTION: Never insert or pull cards or any cable from the equipment with the “power on” under any circumstances. 2