Maintenance 8.2.1 8 Checking flow restrictor Maintenance inter- Every month. val Reason for maintenance A flow restrictor opening at too high pressure will cause a high pressure alarm from the system. In exceptional cases this can result in column rupture. A flow restrictor opening at too low pressure might cause air bubbles in the buffer/solution in the flow path. The position of the flow restrictor: In addition, a flow restrictor is integrated in the mixer. Checking flow re- To check the flow restrictor: strictor Step Action 1 Immerse inlet tubing A1 into deionized water. 2 Start a flow from A1 to injection valve waste by: • Start UNICORN and select System Control: Manual: Flowpath. • Select Injection Valve and Waste. Click Execute. • Select System Control:Manual:Pump. • Select a flow rate of 5 ml/min. Click Execute. 3 Start a flow from A1 to outlet valve waste by: • Select System Control: Manual: Flowpath. • Select Injection Valve and Inject. Click Execute. • p 223