Download User Manual
Transcript
Use 3-6 Loading of Calibrators, Controls or Samples in the Sample Tray ! Take care when opening and closing the reagent drawer: rough handling when opening and closing the drawer may cause the internal reagents to spill. Access authorised (Drawer unlocked). Sampling under way. Access denied (Drawer locked). Any position of the sample tray may be used to install samples, calibrators and controls. Each of these have an associated icon shown in the table below: Calibrator Serum, Plasma, Other Sample Urine Control 3-6-1 Installation using a barcode • If the light is green, open the drawer. Calibrators and controls Scan the barcode with the reader located on the front face. The analyser will beep when the barcode has been correctly read. • Within 5 seconds, place in a free position on the rack. If the barcode is recognised, the analyser will beep a second time. On the interface, the display will indicate that the position has been identified. If no second beep is heard, the position is considered to be occupied but not identified. In this case, re-start the barcode reading and installation process. • • • • • • If the barcode label is illegible, click on Use the keyboard to enter the barcode identifier. Select ‘Vial association with samples tray’ and click ‘OK’. Click on ‘Configure’. The position is configured. ! . For the calibrators and controls supplied by IDS, enter the 12 figures from the barcode: the product, the lot number, the expiry date as well as the table of corresponding values are automatically entered. IDS-iSYS User Manual - Rev C1 Software version V4.XX Use 3-6 48