Maintenance
5-2 Weekly Maintenance (continued)
5-2-1 General Maintenance (continued)
5-2-1-3 Cleaning the rinsing well
WARNING:
RISK OF BIOLOGICAL CONTAMINATION
This part of the analyser is in contact with biological samples. There is, therefore, a
potential risk of infection.
Wear disposable gloves for all handling procedures.
•
•
•
•
Open the lid.
Pour alcohol into the rinsing well.
Use a cotton bud soaked in alcohol to clean the well.
When maintenance is complete, click on the corresponding
maintenance by clicking on YES.
, then confirm the
5-2-1-4 Cleaning the reagent barcode reader mirror
!
•
•
•
•
Do not spray any product onto the medical device.
Open the reagent drawer.
Clean the mirror with absorbent paper soaked in alcohol.
Close the reagent drawer again.
When maintenance is complete, click on the corresponding
maintenance by clicking on YES.
, then confirm the
5-2-1-5 Checking dilutors and IDS-iSYS System Liquid pumps
•
•
•
•
•
Remove the cover from the right-hand hand side of the analyser.
Check for any leaks on the tubing connections at the outlet of each dilutor.
Check for any leaks on the IDS-iSYS System Liquid pumps.
Replace the cover.
When maintenance is complete, click on the corresponding
maintenance by clicking on YES.
IDS-iSYS User Manual - Rev C1
Software version V4.XX
, then confirm the
Maintenance 5-2