Quality Assurance
Logs
◆
Every time that an analysis made on a QC is rejected, a notification is done.
5.4. Reagent logs
◆
For every reagent change, a notification is done.
5.5. Settings logs
◆
At every modification in the settings, a notification is done.
5.6. Maintenance logs
◆
For every maintenance, a notification shall be done by the user or technician in this screen.
After any maintenance operation, open the «Maintenance» logs screen. Press the «Insert»
key in order to enter an action, the duration and a comment about this action. Press «ok»
key to save the new entry.
5.7. Error logs
◆
A notification shall be done when an alarm is displayed
5.8. Host logs
◆
Each time a file from the Host is rejected, a notification is done. This rejection is generated
in the software rather than the raster when, for example:
– An order is pending a ReRun
– The rack on which the tube is found is in progress of an analysis
5.9. Blank cycle logs
For every Startup analysis performed, a notification is done.
◆
A Startup has passed if the cycle is launched and the WBC, RBC, HGB and PLT parameters
are within acceptable limits. the notification line is displayed in green.
◆
A Startup has failed if the cycle is launched and if the three consecutive analyses done
gave wrong results on the WBC or RBC or HGB or PLT parameters, the notification line is red.
◆
A Startup has failed if HGB blank value is not within acceptable limits. The notification
line is red.
◆
The default limit values are:
– WBC = 0.3 103/ mm3
– RBC = 0.03 106/mm3
– HGB = 0.3 g/dl
– PLT = 7 103/mm3
Pentra 80 - User Manual - RAB108EA