Download Manual - Astronics Test Systems
Transcript
Publication No. 980673-055 Rev. A Sequence of Events 1260-22 User Manual When a reset event occurs, the following sequence takes place: • The 1260-22 recognizes the reset event and sets a logic latch. • All relays are opened on the 1260-22. This occurs within the time it actually takes the contacts to open, typically fifteen milliseconds. • If the scope of the reset function is set to global, then all relays on all switch products connected to the same Option01T are also opened. This happens independently of software. • An interrupt is generated to notify the Option-01T that a reset event has occurred. This informs the Option-01T that the relays have opened, so that it remains up-to-date on the relay status. • The relays remain open and do not respond to any commands from the Option-01T. • After the reset condition is removed (the external switch is returned to its normal position), the relays remain open until otherwise commanded by the Option-01T. Once a reset condition has occurred, no communications can occur between the Option-01T and any switch modules until the reset condition is removed. When the reset condition is removed, the relays remain in the open state until operated under software control. If the user issues a relay command while a reset condition is applied, then the following message shall be received: -300 “DEVICE-SPECIFIC ERROR: EXTERNAL RESET PREVENTS EXECUTING OF COMMAND OR OPERATION.” Power and Module Cooling Considerations Astronics Test Systems The 1260-22 is a high density, high power switch module capable of switching high current levels. Because of this certain precautions should be applied when using the switch module. Module Operation 3-11