Download User`s Manual
Transcript
<3 INSTALLATION AND WIRING> 3-1 3 Installation and Wiring 3.1 Mounting of the Analyzer Refer to the SAFETY section (Safe handling and relocation) for instructions on holding and moving the analyzer. The TDLS220 is suitable for wall mounting by means of the four corner located mounting holes as shown below: The holed are ½” sized for typically 3/8” or ¼” bolts and should be used with appropriate sized flat and locking washers. Ensure the analyzer is securing fastened and that there is suitable access for maintenance, etc. The analyzer is designed for operation outside buildings, at normal environmental conditions (see Environmental specifications, page 11). However, it must be protected from direct rain and snow fall. Mounting in a vibration free environment will ensure prolonged service life. Mounting in vibration prone areas may introduce operational issues. Please consider mounting location with care. The actual measurement is NOT affected by vibration, it is an instrument life-time reliability consideration. 22” 22” 22” Figure 1 - Mounting Dimensions IM 11Y01B02-11E-A 3rd Edition :August 14, 2012-00