To test the tipping bucket, pour water in to the tipping and observe the read-out. The read-out will be updated rapidly. When you are done, press the > (DONE) key, and then select CANCEL from the confirmation screen to throw away the test value. To correct the tipping bucket, press * (SET), and enter the new desired level in to the following screen: TIPS=+00000005.25 <OK> <CANCEL> Select OK to accept the change and return to the live display. Press the > (DONE) key, and then select OK from the confirmation screen to accept the new value. Bringing the Benefits of Real-Time Data Collection to the World Sutron Corporation, Tel: 703-406-2800, http://www.sutron.com 195