Download Operating Manual

Transcript
Using Swinging-Bucket Rotors
Rotor Preparation and Loading
3
Remove the bucket gaskets or O-rings and coat them lightly but evenly with silicone vacuum
grease (335148).
a. Install gaskets or O-rings in the buckets.
Cap
Gasket
Bucket
CAUTION
Never run a filled bucket without a gasket or O-ring, as the bucket contents may
be lost, leading to rotor imbalance and possible failure.
4
Dry the exterior of the tubes.
• (Moisture between the tube and the bucket may lead to tube collapse and increase the force
required to extract the tube.)
a. Slide the filled and sealed tubes into the buckets.
•
5
Loaded buckets can be supported in the bucket holder rack available for each rotor.
Use the required spacers and/or floating spacers, if necessary, to complete the loading
operation.
a. If OptiSeal tubes are being used, install a spacer over each plugged tube (refer to the
applicable rotor manual).
1) Leave buckets without tubes completely empty.
Spacer
Tube
Plug
Tube
LR-IM-24AC
5-5
5