4. Loading the Tip Holder Assembly onto the Processor SP
a) The Tip Holder Assembly is a plastic holder that contains two Pipette Tips and a rubber Tip Seal.
Each Pipette Tip contains a filter and an O-ring on top.
Pipette Tip
O-Ring
Tip Seal
b) Before using the Tip Holder Assembly, check the top of each Pipette Tip for the O-ring and confirm
that the rubber Tip Seal is sitting straight and flush between the tips. If either is missing, replace with
a new Tip Holder Assembly.
c) Insert the Tip Holder Assembly into the Drawer Assembly. The image below shows a properly loaded
Tip Assembly. The Tip Holder Assembly can only be loaded in one location and orientation in the
Drawer Assembly. For orientation, there are two holes on the deck of the Drawer Assembly that fit
each Pipette Tip and the opening to the Tip Seal should face away from Processor SP.
Tip Holder
Assembly
Page 10 of 59
®
Verigene Respiratory Pathogens Flex Nucleic Acid Test (RP Flex)
027-00050-01, Rev. B; September 2015