Download VME-MXI-2 User Manual - National Instruments
Transcript
Chapter 3 VME-MXI-2 Configuration and Installation MXIbus Termination The first and last MXIbus devices connected to the MXIbus—whether it is a single MXI-2 cable or daisy-chained MXI-2 cables—must terminate the MXIbus. Any MXIbus devices in the middle of a MXIbus daisy chain must not terminate the MXIbus. The VME-MXI-2 automatically senses if it is at either end of the MXIbus cable to terminate the MXIbus. You can manually control MXIbus termination by defeating the automatic circuitry. Use switches 3 and 4 of the four-position switch at location U21 to control whether MXIbus termination is automatic (Figure 3-4a), on (Figure 3-4b), or off (Figure 3-4c). The settings of switches 1 and 2 have no effect on MXIbus termination. Use switch 3 to select whether you want the VME-MXI-2 to automatically control termination of the MXIbus. Switch 4 lets you manually control whether to terminate the MXIbus when automatic termination is turned off. Switch 4 has no effect when switch 3 is set for automatic MXIbus termination; you must turn off automatic termination if you want to manually control termination. VME-MXI-2 User Manual 3-6 © National Instruments Corporation