Download GDS-800-programmer_manual
Transcript
GDS-806/810/820/840 Programming Manual 1. Introduction Users can drive this digital storage oscilloscope by using the GPIB (General Purpose Interface Bus) system with a computer, or from a computer across the RS-232 serial connection. Commands sent over either interface can read or set any GDS-806/810/820/840’s instructions. This chapter explains how to carry out the following tasks. Notes for GPIB installation If you are setting up the oscilloscope with a GPIB system, please check the following regulations: z Only a maximum of 15 devices can be connected to a single GPIB bus. z Do not use more than 20 m of cable to connect devices to a bus. z Connect one device for every 2 m of cable used. z Each device on the bus needs a unique device address. No two devices can share the same device address. z Turn on at least two-thirds of the devices on the GPIB system while you use the system. z Do not use loop or parallel structure for the topology of GPIB system. 2