Download Operator`s manual
Transcript
Alarms, Advisories, and Prompts Use this troubleshooting information in conjunction with appropriate hospital procedures. Responding to an advisory, alarm, or fault message 1. Press QUIET. • Audio tone stops. • Red light flashes on affected channel. 2. Press affected channel A , B or C . • Alarm Information page is displayed. 3. Take appropriate action(s) indicated on the display. 4. Press START STOP to resume infusion. • Channel starts infusing. 5. Press STANDARD DISPLAY OR • Display reverts to Standard Display after one minute. 6. Verify settings. 7. Verify flow. Directions for Use MedSystem III® infusion pump Models 2865/2866 Alarms, Advisories, and Prompts 45