Download Micros 60 CS-CT - Helena Laboratories
Transcript
Micros 60 1.3. Service Functions Several service functions are available for the operator to clean and check the instrument These functions are accessable from the Main Menu. Use the “Down “ arrow to select 4 SERVICE. Press the “ENTER” key to access these sub-menus as indicated on the display: SERVICE 12:20 > 1 - BACKFLUSH 2 - DRAIN CHAMBERS Backflush This service function will allow the operator to clean the counting chamber apertures in case of blockage. From the Service Menu, select 1 BACKFLUSH, then press the “ENTER” key to start the process. This cycle takes approximately (20) seconds to complete. • Open the instrument door and verify that liquid is being pulled through the apertures. (watch for few micro-bubbles in the count tubing, coming from the count head). • When the backflushing starts, verify that liquid is being pushed back through the apertures. (watch for few micro-bubbles coming into the chambers through the count head). • If you cannot see any micro-bubbles moving through tubing or in the chambers, the aperture may still be blocked. In this case, perform a Concentrated Cleaning. Drain Chambers This service function will allow the operator to check for proper chamber draining and to maintain some of the parts of the hydralic manifold as it flushes waste out of the instrument. • This cycle takes approximately (7) seconds to complete. From the Service Menu, select 2 DRAIN CHAMBERS, then press the “ENTER” key to darin the chambers. 4 • Verify the proper draining of the waste through the waste tubing located on the back of the instrument. if it is not flushing waste sufficently, check the waste tubing connection and verify that there is no blockage or crimping of the tubing. • Verify that both the WBC and RBC chamber are draining properly, not slow but fairly rapid.If not, view the troubleshooting section. Prime Reagents This service function will allow the operator to prime reagents either individually or all at once, depending on the type of system you have. (Separate Reagents or Minipak). • From the Service Menu, select 3 - PRIME REAGENTS. There are (2) types of ABX MICROS 60 systems: 1 - BOTTLE SYSTEM (Separate reagents). The prime reagents sub-menu will indicate: 1 - ALL REAGENTS 2 - DILUENT 3 - LYSE 4 - CLEANER • Press the “ENTER” key to access this submenu for bottled units. Select the reagent or reagents to be primed, then press the “ENTER” key to start the prime cycle. 2 - PACK SYSTEM (ABX Minipak-all reagents contained in (1) pack). Diluent ,Lyse, and Cleaner. The Prime reagents sub-menu will indicate: 1 - CHANGE PACK 2 - CAPACITY 3 - PRIME PACK • Press the “ENTER” key to access this submenu. Select either Changing the Pack if the present pack is low, or Priming the current pack if the cycles are sufficient. Step by Step instructions will be given on the display when installing a Minipak. RAB 043 CAS /section 6: Maintenance and Troubleshooting