Download downloaded (in PDF

Transcript
Section 4: Front Panel Operations
Model 2100 6 1/2-Digit Resolution Digital Multimeter User’s Manual
Figure 4-6
Front panel auto-range
CAT
CAT
1000V
600V
INPUT
V
SENSE
4 WIRE
HI
200V
PEA K
FUNCT ION
PREV
DCI
ACI
4
PERIOD
DCV
AC V
2
FREQ
DISPLAY
FILTER
NEXT
OFF
ON
POWER
SETUP
DIGITS
EXTRIG
HOLD
SINGL E
AUTO
TRIGGER
TRIGGER
CONT
TEMP
PREV
LIMIT S
MX+B
dBm
dB
LOCK
RATIO
%
MIN/MA X
NULL
ESC
3A
NEXT
INPUTS
F
ENTER
RECA L L
SHIFT
CONFIG
A UTO
500V
PEA K
RMS
R
PREV
A UTO
FRONT/REAR
L OCA L
1000V
PEA K
LO
RANGE
MATH
STORE
RATIO
REF
3A 250V
AMP S
MENU
NEXT
MEMORY
To set the range from a remote interface:
Use the following commands from your PC terminal to set the range:
CONFigure:<function> <range>,<resolution>
MEASure:<function>? <range>,<resolution>
SENSe:<function>:RANGe <range>
SENSe:<function>:RANGe:AUTO {OFF|ON}
Rate (integration time)
Definition
Integration time is the period of time the analog-to-digital (A/D) converter takes to sample the input
signal. The integration time feature allows you to optimize either the speed or the resolution, as
well as the noise rejection and accuracy of the measurement. The unit of the integration time is in
PLC (power line cycles); one PLC for 60Hz is 16.67ms, and for 50Hz is 20ms.
The Model 2100 has four different integration times you can select: 0.02 PLC, 0.1 PLC, 1 PLC,
and 10 PLC.
Default
For DCV, DCI, and resistance measurement, the default integration time is 1 PLC. The user
selected value is stored in volatile memory, and the default value will be restored when the meter
is turned-off.
How to set the integration time:
You can set the integration time either through the front panel or the remote interface operation.
To set the integration time from the front panel:
Integration time is set indirectly when you select the measurement resolution. Please refer to
“Resolution setting (digits)” later in this section for details about how to set resolution or digits.
Table 4-1 shows the relationship between the resolution and the integration time.
To set the resolution for your measurement from a remote interface:
Use the following commands on your PC terminal to set the resolution:
4-10
Return to Section Topics
2100-900-01 Rev. B / July 2007